Product Center
Humidity Calibrator
Model: JT8040
33%~85%RH 6 kinds of saturated salt solutions, providing 6 different relative humidity calibration points.
The portable humidity calibrator uses the principle that the saturated salt solution produces a fixed relative humidity value in the contact surface with the air under stable conditions, and provides a faster, more accurate, economical and portable humidity calibration method. It is an ideal choice for users such as humidity sensors, portable humidity test instruments, and metering tests.
Calibration principle: at a certain temperature, the partial pressure of water vapor above the saturated salt solution is in equilibrium with the environment and is a constant Basis: JJG 2046-1990 Humidity Measuring Instrument Verification System The humidity value of the saturated salt tank is in line with the value stated in the "Standard Relative Humidity Table of Saturated Salt Solution" by the International Legal Organization (OIM L):
℃\# | 7#?Potassium acetate | 9#?magnesium chloride | 11#?potassium carbonate | 13#?Sodium Bromide | 17#?sodium chloride | 19#?potassium chloride |
10 | 23.38±0.53 | 33.47±0.24 | 43.14±0.39 | 62.15±0.60 | 75.67±0.22 | 86.77±0.39 |
15 | 23.40±0.32 | 33.30±0.21 | 43.15±0.33 | 60.68±0.51 | 75.61±0.18 | 85.92±0.33 |
20 | 23.11±0.25 | 33.07±0.18 | 43.16±0.33 | 59.14±0.44 | 75.47±0.14 | 85.11±0.29 |
25 | 22.51±0.32 | 32.78±0.16 | 43.16±0.39 | 57.57±0.40 | 75.29±0.13 | 84.34±0.26 |
30 | 21.61±0.52 | 32.44±0.14 | 43.17±0.50 | 56.03±0.38 | 75.09±0.11 | 83.62±0.25 |
35 | ? | 32.05±0.13 | ? | 54.55±0.38 | 74.87±0.12 | 82.95±0.25 |
?
6 kinds of saturated salt solution options
JT8040-23/33/43/59/75/85 provides 6 kinds of saturated salt solution options, from 23% to 85%, suitable for humidity probes with a diameter of 13~16mm
Basic technical parameters of JT8040-XX series humidity calibrator (saturated salt tank)